Evaluation of calibration uncertainty reference conductivity cell of calibrator СT01«Idronaut» «Installations for measuring conductivity of sea water VNIRO»
https://doi.org/10.36038/2307-3497-2021-184-99-111
Abstract
The results of analysis are given for the experimental studies of the metrological characteristics of the reference conductivity cells carried out during the calibration of the conductivity sensors of CTD probes from 2008 to 2017 using “Installation for measuring conductivity sea water VNIRO” The factors determining the uncertainty of calibration of the conductivity reference cell are considered. The calculations were carried out for the combined standard uncertainty (CSU) of the cell calibration and standard uncertainties of its components. The expanded CSU of their calibration at any point in the range (53÷27) mS/cm was concluded not to exceed (2.2÷1.0)·10–3 mS/cm, which corresponds to the relative expanded CSU calibration equal to 4·10–5. The analysis of the contribution of individual components to the total uncertainty is given. It was shown that CSU of the cell calibration is mainly estimated by the CSU of determining the reference value of the CSU SSW, due to the uncertainty in determining of the sample temperature inside the cell because of instability and inhomogeneity of the temperature field around it uΣС(Т) . The value of uΣС(Т) varies from 9.7·10–4 mS/cm to 3.6·10–4 mS/cm at the level of instability of sea water in the thermal tank ranging from (1.1 ± 0.17)·10–3 °C at a temperature of 25 °C to (0.53±0.19)·10–3 °C at a temperature of 1 °C. The drift of the static cell conversion function for 413 days did not exceed ±0.003 mS/cm.
About the Author
A. N. RamazinRussian Federation
Moscow
References
1. Levashov D.E. 2003. Tekhnika ekspedicionnyh issledovanij: Instrumental’nye metody i tekhnicheskie sredstva ocenki promyslovo-znachimyh faktorov sredy [Facilities and technique for marine surveys: Instrumental methods and components for estimating the fishing-significant characteristics of water medium]. M.: Izd-vo VNIRO. 400 s.
2. Ramazin A.N. 2016. Nekotorye nauchno-metodicheskie aspekty metrologicheskogo obespecheniya izmerenij elektroprovodnosti i temperatury morskoj vody s pomoshch’yu STD-zondov [Some scientific and methodological aspects of metrological supply of conductivity and sea water temperature measurements with CTD-probes] // Trudy VNIRO, T. 159, S. 167–182.
3. Ramazin A .N., Levashov D.E., 2016. Metodicheskie rekomendacii po kalibrovke i metrologicheskomu obespecheniyu izmeritelej parametrov vodnoj sredy [Guidelines for calibration and metrological support of measuring instruments for the parameters of the aquatic environment]. M.: Izd-vo VNIRO. 127 s.
4. Ramazin A.N. 2018. Ocenka neopredelyonnosti kalibrovki i izmerenij datchikov temperatury STD-zondov «Sea Bird» [Evaluation of calibration and measurement uncertainty of temperature sensors of CTD-probe “Sea Bird”] //Sistemy kontrolya okruzhayushchej sredy. Sevastopol’: IPTS. 2018. Vyp.14 (34), S.33–42.
5. Rukovodstvo po vyrazheniyu neopredelennosti izmereniya. 1999.Perevod s angl. pod nauch. red. prof. Slaeva V.A. SPb.: VNIIM im. D.I. Mendeleeva. 134 s. (Guide to the Expression of Uncertainty in Measurement. 1993. First edition. ISO, Switzerland. 101 pp.)
6. Bacon, S.; Culkin, F.; Higgs, N.; Ridout, P. 2007. IAPSO Standard Seawater: definition of the uncertainty in the calibration procedure, and stability of recent batches // J. of Atmospheric and Oceanic Technology. 24 (10). pp.1785–1799. Accessible via: https://journals.ametsoc.org/doi/pdf/10.1175/JTECH2081.1 20.08.2019.
7. Brinkmann F, Dam N.E., Deak E., Durbiano F., Ferrara E., Fuko J., Jensen H.D., Mariassy M., Shreiner R.H., Spitzer P., Sudmeier U., Surdu M., Vyskocˇil, Vyskocˇil L. Primary methods for the measurement of electrolytic conductivity. Accessible via: http://www.researchgate.net/publication/227274619. 20.08.2019.
8. Culkin F., Smith N.D. 1980. Determination of the Concentration of Potassium Chloride Solution Having the Same Electrical Conductivity, at 15 °C and Infinite Frequency, as Standard Seawater of Salinity 35.0000‰ (Chlorinity 19.37394 ‰) // IEEE J. of Oceanic Engineering. V. OE-5. N 1, p. 22–23.
9. IOC, SCOR and IAPSO. 2010. The international thermodynamic equation of seawater — 2010: Calculation and use of thermodynamic properties. Intergovernmental Oceanographic Commission, Manuals and Guides. No. 56, UNESCO (English), 196 pp. Accessible via: http://www.teos-10.org/TEOS‑10_Manual.pdf. 20.08.2019.
10. Lewis E.L., Perkin R.G. 1981. The Practical Salinity Scale 1978: conversion of existing data // Deep-Sea Res. V. 28a. № 4. pp. 307–328.
11. Mangum B.W., Furukawa, G.T. 1990. Guidelines for Realizing the International Temperature of 1990 (ITS-90). NIST Tech. Note 1265. 190 p.
12. Poisson A. 1975. Measurement of absolute electrical conductivity of standard sea water on the basis of KCl as standard // UNESCO Technical Paper in Marine Science. Annex 3, vol. 24–61 p.
13. Perkin R.G., Lewis E.L. 1980. The practical salinity scale1978: Fitting the data // IEEE J. Oceanic Engineering.V. OE5. no. 1, pр. 9–16.
14. Seitz S, Spitzer P. Traceable measurements of conductivity and its relation to salinity. Working Group for Electrochemistry. Physikalisch-Technische Bundesanstalt. 31 р. Accessible via: https://www.ptb.de/cms/fileadmin/internet/fachabteilungen/abteilung_3/3.1/3.13/230ptbsem/230ptbsem_ptb-seitz.pdf/ 20.08.2019.
15. Stueben D., Haushahn P., Stueben K. 1994. MINIBAT — a new, simple system for in-situ measurement, mapping and sampling of dissolved trace elements in aquatic systems // Int. underwater syst. desing. V. 16. N 5. рр. 5–14.
16. Shreiner R.H., Pratt K.W. 2004. Standard Reference Materials: Primary Standards and Standard Reference Materials for Electrolytic Conductivity. NIST Special Publication 260–142. 26 p.
17. Yerofeyev P.N., Ponomareva L.S., Ramazin A.N., 1977. Criteria for the definition of salinitywaterby electroconductivity with corrections for temperature and pressure. Four Russian papers on salinity. Inst. of Ocean Sciences, Patricia Bay. Victoria, B.C., рр. 1–56. Accessible via: http://www.dfo-mpo.gc.ca/Library/54947.pdf. 20.08.2019.
Review
For citations:
Ramazin A.N. Evaluation of calibration uncertainty reference conductivity cell of calibrator СT01«Idronaut» «Installations for measuring conductivity of sea water VNIRO». Trudy VNIRO. 2021;184:99-111. (In Russ.) https://doi.org/10.36038/2307-3497-2021-184-99-111